Dr. Fritz | Veterinary Endoscopy | Catalog 2022

109 E 109 Dr. Fritz Endoscopes www.dr-fritz.com . dr.fritz@dr-fritz.de Europe-Germany 2022 108 Dr. Fritz Endoscopes www.dr-fritz.com . dr.fritz@dr-fritz.de Europe-Germany 2022 Cleaning recommendations for field practice During main reprocessing please follow the instructions of the manufacturers! Equine Dynamic and Overground Endoscopy ETL III, Equine Training Laryngoscope – the latest generation – Intermediate cleaning after every use during the day. Full cleaning and disinfection is mandatory in the evening! 2. Clean scope outside with damp cloth and cleaner (Dr. Fritz enzym cleaner FLEX), diluted as instructed, moisten. 3. Make sure not to forget distal end and front lens 4. Fill syringe with cleaning solution 5. Flush channels (A) via air/water valve and (B) the working channel 4. 6. Wipe scope outside with damp cloth see no.2 7. 7. Fill syringe with alcohol solution 8. Flush channels with alcohol solution (see no.5). Afterwards fill syringe with air and press it through the channels to dry and to remove alcohol. 9. For transportation store your scope within a protection cloth or film. Use a cotton tissue gaze for the distal tip of the endoscope to collect rest water We highly recommend to use gloves while examining your patients to reduce contamination of your scope. ☛ ☛ 1. Preparations: 1 Dr. Fritz enzymatic cleaning FLEX 2 alcohol 3 syringe 4 gloves 5 protection cloth 1. 1 2 3 4 5 2. A B 5. . 3. 6. 8. 9.